Principle:
The immunolabeling of slides, coverslips or free-floating sections is a time consuming process. On the bench the whole process will take a few hours or days. Immunolabeling can be direct, indirect or a sandwich assay and can use fluorescent or enzyme based chromogens. Various blocking steps are employed and numerous rinses are incorporated in the overall process. Many steps and different reagents are therefore being used in a typical immunoassay.
Microwave-assisted immunolabeling has demonstrated the ability to greatly shorten turn-around times regardless of the labeling protocol used. This statement has been validated through numerous workshops and recent publications. The ability to use microwave radiation effectively has evolved through better control of the cavity environment during microwave exposure. Slow controlled sample heating, on the order of 2-3°C a minute, are the norm.
How to Begin:
The starting sample can be a free-floating frozen or Vibratome section, paraffin section, tissue culture plate or small embryo.
Starting Point: Paraffin section on a glass slide.

Microwave-assisted immunolabeling of human skin with a universal strepavidin/biotin immunostaining system. A polyclonal cytokeratin wide spectrum primary antibody was used along with AEC as the chromagen. Total staining time <60 minutes in microwave.
Starting Point: Free floating section/tissue.
|
Retinas were fixed in 4% paraformaldehyde in 0.1M phosphate buffer (pH 7.4) overnight at 4°C. Following labeling the tissue was rinsed 6 x 20 minutes in buffer prior to beginning antibody labeling. The bench labeling protocol requires 7 days. These labeling results were completed in an afternoon using microwave-enhanced labeling (PELCO BioWave® with SteadyTemp) during a workshop held at the Univ. of Minnesota Imaging Center (Mark Sanders, Director - May 17-19, 2006). The retinas were double-labeled for:
Primary antibody labeling was done at 170W for 12 minutes (4 on - 4 off - 4 on) under vacuum (15" Hg). Secondary antibody labeling was done at 170W for 6 minutes (2 on - 2 off - 2 on) under vacuum (15" Hg). Images were collected on a Nikon C1si Confocal Microscope. |
![]() |
|
Retinas were fixed in 4% paraformaldehyde in 0.1M phosphate buffer (pH 7.4) overnight at 4°C. Following fixation the tissue was rinsed 6 x 20 minutes in buffer prior to beginning antibody labeling. The bench staining protocol required 7 days. The labeling results were completed in an afternoon using microwave-enhanced labeling during a workshop held at the Univ. of Minnesota Imaging Center (Mark Sanders, Director - May 17-19, 2006). The retinas were triple-labeled for:
Primary antibody labeling was done at 170W for 12 minutes (4 on - 4 off - 4 on) under vacuum (15" Hg). Secondary antibody labeling was done at 170W for 6 minutes (2 on - 2 off - 2 on) under vacuum (15" Hg). Images were collected on a Nikon C1si Confocal Microscope. |
![]() |
| Microwave Immunolabeling Protocol: Start at the step that is appropriate for the labeling protocol being used. |
|||
| No. | Step | Microwave Settings Wattage/Temp Restriction |
Time |
| 1. | Coplin Jar - 50 ml. xylene - deparaffinization | 250 W | 4 min. |
| 2. | Coplin Jar - 50 ml. 95% ethanol - rehydrate | 250 W | 1 min. |
| 3. | Coplin Jar - Wash in tap water - rehydrate | 30 sec. | |
| 4. | Antigen Retrieval (if required) | No specific method is recommended | |
| 5. |
Transfer Slides to Coverplate System1 |
||
| 6. |
Block Endogenous Peroxidase - 3 drops (~100µl) |
150-250 W | 1 min. |
| 7. | Buffer Rinse - ~1 ml (1 to 3 x 1 min.) | 150-250 W | 1 min. |
| 8. |
Blocking Step - 3 drops to ~1 ml |
150-250 W | 3 min.2 |
| 9. | Primary Antibody - 3 drops (~100µl) Vacuum + increased time for thick tissues (>20µm) |
150-250 W 15-20" Hg |
6 min.3 12 min.4 |
| 10. | Buffer Rinse - ~1 ml (1 to 3 x 1 min.) | 150-250 W | 1 min. |
| 11. | Secondary Antibody - 3 drops (~100µl) Vacuum + increased time for thick tissues (>20µm) |
150-250 W 15-20" Hg |
6 min.3 12 min.4 |
| 12. | Buffer Rinse - ~1 ml (1 to 3 x 1 min.) | 150-250 W | 1 min. |
| 13. | Tertiary Attachment - 3 drops (~100µl) Vacuum + increased time for thick tissues (>20µm) |
150-250 W 15-20" Hg |
6 min.3 12 min.4 |
| 14. | Buffer Rinse - ~1 ml (1 to 3 x 1 min.) | 150-250 W | 1 min. |
| 15. | Chromagen - 100-500µl | 150-250 W | 1-6 min. |
| 16. | Rinse in D.I. water - ~1 ml (1 to 3 x 1 min.) | 150-250 W | 1 min. |
| 17. | Counter Stain - 100-500µl | 150-250 W | 1-3 min. |
| 18. | Rinse in D.I. water - ~1 ml (1 to 3 x 1 min.) | 150-250 W | 1 min. |
| 19. | Remove Slides from Coverplate System | ||
| 20. | Mount Coverslip and View | ||
1The Coverplate System is described in Fig. 1
23 min. corresponds to a preprogrammed time sequence (1 min. on-1 min. off-1 min. on)
36 min. corresponds to a preprogrammed time sequence (2 min. on-2 min. off-2 min. on)
412 min. corresponds to a preprogrammed time sequence (4 min. on-4 min. off-4 min. on)
|
Figure 1. A. A front view of the coverplate that is placed on the glass slide to form a capillary gap for the staining reagents to flow through. The arrow indicates the notch in the front clip of the coverplate that connects with the tab in Fig. 1C (arrow). B. The rear view of the coverplate showing the tabs (<) that position and hold the slide (tissue side down) in position prior to placing the coverplate and slide into the cassette base. Place approximately 2ml of buffer on the coverplate prior to placement of the slide. Check to see that the capillary gap is completely filled with fluid (no bubbles). If bubbles are present rewet the surface and repeat slide placement. C. The cassette base holds the coverplate and slide complex together to maintain the capillary gap during microwave. The lid is not needed for microwave protocols
|
![]() |
The PELCO BioWave® Pro is newest model in our long history of microwave development. It has features that are unique to previous models and ideally suited for immunolabeling applications of all types. These features include:
| State-of-the-art control is demonstrated by the degree of heating and uniformity of temperature between 2 types of wellplates (6 and 12-well) exposed to two different continuous wattage outputs for 1 minute. |
||||||||||
|
630W of continuous power for 1 minute + 3x800ml water loads positioned in the microwave cavity to create a cold spot.
Power output and heating profile typical of a standard laboratory microwave oven. |
![]() |
|||||||||
|
212W of continuous power for 1 minute + Patented PELCO ColdSpot® placed in the microwave cavity to create a large area of uniformity.
Power output and heating profile typical of the PELCO BioWave® Pro set to the parameters best suited to immunolabeling . |
![]() |
|||||||||