Specifications |
|
| Chamber size: | 120mm OD x 120mm high (4.75 x 4.75") |
| Target: | Au fitted as standard, Au:Pd, Pt, Pt:Pd optional. 57mm dia. x 0.1mm thick |
| Sample table: | Holds 12 SEM pin mounts Height adjustable through 60mm |
| Sputter supply: | Microprocessor based, safety interlock, variable, 40mA maximum |
| Sputter head: | Low voltage planar magnetron. Quick target change Wrap-around dark-space shield |
| Analog metering: | Vacuum Atm - 0.001mb |
| Thickness monitoring: | Model MTM-10 (optional) |
| Control method: | Manual gas and current control Digital timer (0-300 seconds) with pause Manual vent |
| Desktop system: | Vacuum pump is mounted on desktop compatible anti-vibration table All metal integrated coupling system |
Requirements |
|
| Electrical: | 100-120 or 200-240 VAC, 50/60Hz (specify on order) |
| Power: | 300VA (coater and pumping system #7010 together) |
| Argon gas: | Purity, min. 99.99% Working pressure 5-6 psi (0.4 bar) Hose connection, 6.0mm (1/4") |

|
1. Control unit 2. Vacuum chamber 3. Thickness monitor |
4. Rotary pump 5. Anti-vibration base |